Virucidal efficacy of a sonicated hydrogen peroxide system (trophon EPR) following European and German test methods.

Becker, Britta
Bischoff, Birte
Brill, Florian H H
Steinmann, Eike
Steinmann, Jochen
Citations
Altmetric:
Advisors
Editors
Other Contributors
Issue Date
2017-01-01
Submitted date
Other Titles
Abstract
The virucidal efficacy of an automated ultrasound probe disinfector (trophon® EPR) was evaluated in a three step procedure according to European and German test methods. This system uses sonicated hydrogen peroxide mist (35%) at elevated temperature (50°C) in a closed chamber with control of all parameters within a 7 minute cycle. Methods: In the first step of examination, the peroxide solution was tested in a quantitative suspension assay according to the Guideline of Deutsche Vereinigung zur Bekämpfung der Viruskrankheiten (DVV) e.V. and Robert Koch-Institute (RKI) and in parallel with the European Norm EN 14476 with all test viruses creating a virucidal claim. In the second step, the virucidal efficacy of the hydrogen peroxide solution was evaluated in a hard surface carrier test according to the Guideline of DVV with adenovirus, murine norovirus and parvovirus simulating practical conditions. Finally, the efficacy was evaluated by the automated system using stainless steel carriers inoculated with test virus and positioned at different levels inside the chamber.
Citation
Publisher
Journal
PubMed ID
PubMed Central ID
Additional Links
Embedded video
Type
Article
Language
Description
Series/Report no.
ISSN
2196-5226
EISSN
ISBN
ISMN
Gov't Doc #
Sponsors
License
Attribution-NonCommercial-ShareAlike 3.0 United States