Loading...
Thumbnail Image
Publication

INTEGRATED ENZYME THIN FILM METAL ECTRODES

Hintsche, Rainer
Möller, B.
Dransfeld, I.
Scheller, Frieder
Citations
Altmetric:
Advisors
Editors
Other Contributors
Issue Date
1992
Submitted date
2024-04-10
Other Titles
Abstract
Miniaturized electrodes have been constructed as transducers of biosensors. Onto oxidized silicon wafers Ti as adhesion and Au as conductivity layers have been deposited. A 120nm Pt layer forms the upper layer for amperometric and impedimetric electrodes, whereas Sb/SbO or Ag/AgCl composite layers are the electro-chemically active layers of the potentiometric and the reference electrodes, respectively. They can be combined in different shapes and sizes at the chips and applied in test strips or flow through devices. The amperometric electrodes with a upper Pt layer are stable at potentials of water electrolysis. The reference Ag/AgCl composite electrode is more resistent against light and dissolution than the common surface bound AgCl layers at silver.The pH dependency of the Sb/SbOelectrode is 56 + 0.1 mV/pH and was observed to be stable over a period of about 30 days, which should be useful for the normal working time of biosensors.
Citation
Biosensors : fundamentals, technologies and applications, 61 - 63
DOI
PubMed ID
PubMed Central ID
Additional Links
Embedded video
Type
Book chapter
conference paper
Language
en
Description
Series/Report no.
GBF monographs ; Volume 17
ISSN
0930-4320
EISSN
ISBN
3527284370
1560812206
ISMN
Gov't Doc #
Sponsors
License
Attribution-NonCommercial-ShareAlike 4.0 International