DEVELOPMENT OF AN ENZYMATIC MULTICHANNEL FLOW INJECTION ANALYSIS SYSTEM FOR MONITORING MAMMALIAN CELL FERMENTATIONS
Van der Pol, J. J. ; Joksch, B. ; Eberhardt, R. ; Biselli, M. ; Wandrey, Ch. ; Tramper, J.
Van der Pol, J. J.
Joksch, B.
Eberhardt, R.
Biselli, M.
Wandrey, Ch.
Tramper, J.
Citations
Altmetric:
Advisors
Editors
Other Contributors
Issue Date
1992
Submitted date
2024-05-08
Files
Loading...
PDF
Adobe PDF, 1.9 MB
Other Titles
Abstract
A multichannel flow injection analysis system described earlier /1/, is adapted to mammalian cell fermentations by increasing the sensitivity of the system. The analysis is based on the conversion of the compounds by immobilized dehydrogenases, resulting in the built up or consumption of NADH which is measured with a fluorescence- detector. The method presented here has a linear glucose determination range of 0.05 g/1 to 3.5 g/1 and a non-linear range for lactate of 0.05 g/1 to 2.7 g/l.
Citation
Biosensors : fundamentals, technologies and applications, 349 - 352