Abstract
A flow-injection system with anodic amperometric detection for glucose determination in human blood serum samples was optimized by a comparison of various arrangements of enzyme immobilization and elimination of interferences such as ascorbic acid, urea and uric acid. The best results were obtained using a compact membrane biosensor in wall-jet flow-through detector, where Pt disc electrode was covered with evaporated Nafion layer, then a polyester membrane with immobilized glucose oxidase and another protective polyester membrane. In the optimized conditions a linear response up to 25mM glucose was observed with detection limit 200uM glucose and sampling rate 120 hr-1 for 20p1 sample volume. Results of flow injection glucose determination in human serum samples were compared with Beckman Glucose Analyser 2 and Kone Dynamic analyser.