Loading...
Thumbnail Image
Publication

ENZYME IMMUNOASSAYS USING FLOW INJECTION ANALYSIS WITH ELECTROCHEMICAL DETECTION

Halsall, H. Brian
Xu, Yan
Heineman, William R.
Citations
Altmetric:
Advisors
Editors
Other Contributors
Issue Date
1991
Submitted date
2024-02-07
Other Titles
Abstract
The utility of FIAEC for enzyme immunoassaysof low detection limit can be constrained bythe existence of a nonfaradaic capacitance current that records with the analytical faradaic signal. Since the magnitudeof this unwanted signal diminishes with smaller potential, conditions can be found where the interference becomes unimportant. This was done here by developing p-aminopheny] phosphate as a substrate for alkaline phosphatase, which yields p-aminophenol as product, oxidizable at between +100 and +300 mV. Additional advantageis gained by matching the compositions of the sample and mobile phase solutions. The useof p-aminopheny] phosphate as a substratefor electrochemical enzyme immunoassay was demonstrated by sensitive assays for theophylline, IgG and alpha-fetoprotein.
Citation
Flow Injection Analysis (FIA) Based On Enzymes Or Antibodies, 169 - 179
DOI
PubMed ID
PubMed Central ID
Additional Links
Embedded video
Type
Book chapter
conference paper
Language
en
Description
Series/Report no.
GBF monographs ; Volume 14
ISSN
0930-4320
EISSN
ISBN
1560811218
3527282491
ISMN
Gov't Doc #
Sponsors
License
Attribution-NonCommercial-ShareAlike 4.0 International