Show full item recordAbstract
The sterilization of potentially infectious animal carcasses is an important biologic safety issue in animal facilities operating as infection or quarantine barriers. However, the literature lacks a validated protocol. Here we describe the validation of an autoclave program suitable for daily use in a small rodent biocontainment unit. We evaluated several procedures for processing mouse carcasses in a standard autoclave. Heat sensors and biologic indicators were implanted inside the peritoneal cavity of dead mice, which were loaded at various densities into IVC cages or metal boxes. Heat sensors revealed broad differences in temperature inside carcasses compared with the autoclave chamber. Achieving the appropriate sterilization temperature was considerably prolonged in carcasses compared with typical laboratory waste material. We show that for 5 cadavers placed well separated inside an IVC, a modified program for mouse cage sterilization using 134 °C for 15 min is suitable. To sterilize approximately 1 kg of carcasses in autoclavable boxes, a period of 6 h is required to reach an effective temperature of 121 °C for 60 min at the center of the waste by using an autoclave program for liquids. In conclusion, we here validated 2 protocols for the sterilization of potentially infectious mouse carcasses, to ensure the application of efficacious procedures.Citation
J Am Assoc Lab Anim Sci. 2019;58(1):87-91. doi:10.30802/AALAS-JAALAS-18-000020.Affiliation
HZI,Helmholtz-Zentrum für Infektionsforschung GmbH, Inhoffenstr. 7,38124 Braunschweig, Germany.PubMed ID
30401066Type
ArticleLanguage
enISSN
1559-6109ae974a485f413a2113503eed53cd6c53
10.30802/AALAS-JAALAS-18-000020
Scopus Count
The following license files are associated with this item:
- Creative Commons
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-ShareAlike 4.0 International
Related articles
- Validating Autoclave Cycles for Carcass Disposal in Animal Biosafety Level 2/3 Containment Laboratories.
- Authors: McGirr R, Sample C, Arwood L, Burch J, Alderman S
- Issue date: 2019 Sep 1
- Thermal Inactivation of Carcasses of Mice and Rabbits Infected with Pathogens of Risk Groups Two to Four.
- Authors: Baldauf HM, Weingartner S, Hofmann K, Mitteregger-Kretzschmar G, Popper B, Bönisch MP, Keppler OT
- Issue date: 2021 Jul 1
- Validation of Autoclave Protocols for Successful Decontamination of Category A Medical Waste Generated from Care of Patients with Serious Communicable Diseases.
- Authors: Garibaldi BT, Reimers M, Ernst N, Bova G, Nowakowski E, Bukowski J, Ellis BC, Smith C, Sauer L, Dionne K, Carroll KC, Maragakis LL, Parrish NM
- Issue date: 2017 Feb
- Measurement of Equilibration Time of Dry Goods Loads in Autoclaves.
- Authors: Shah S, Alayli S, Peng Y
- Issue date: 2019 Jul-Aug
- Technical and Anatomical Considerations for Reproducible Inactivation of Large Animal Carcasses by Steam Sterilization.
- Authors: Schinköthe J, Bartram-Sitzius B, Teifke JP, Pfitzner U, Reiche S
- Issue date: 2021 Mar 1