Show simple item record

dc.contributor.authorWeiss, Thomas
dc.contributor.authorCammann, Karl
dc.date.accessioned2023-08-30T09:47:17Z
dc.date.available2023-08-30T09:47:17Z
dc.date.issued1987
dc.date.submitted2023-08-30
dc.identifier.citationBiosensors International Workshop 1987, 267 - 268en_US
dc.identifier.isbn0-89573-683-7
dc.identifier.isbn3-527-26801-4
dc.identifier.issn0930-4320
dc.identifier.urihttp://hdl.handle.net/10033/623439
dc.description.abstractAn amperometric needle-type glucose sensor (1) consisting of a silver tube (cathode) and a platinum wire (anode) was developed. The sensor was provided with two enzyme layers (glucose oxidase (GOD) in cellulose acetate, catalase). The destructive influence of H,0, generated in the GOD layer could be minimized and the oxygen dependence of the analytical signal was improved.en_US
dc.language.isoenen_US
dc.publisherGBF - Gesellschaft für Biotechnologische Forschungen_US
dc.relation.ispartofseriesGBF Monographs, Volume 10en_US
dc.rightsAttribution-NonCommercial-ShareAlike 4.0 International*
dc.rights.urihttp://creativecommons.org/licenses/by-nc-sa/4.0/*
dc.titleAN AMPEROMETRIC GLUCOSE SENSOR WITH COMBINED ENZYME LAYERSen_US
dc.typeBook chapteren_US
dc.typeconference paperen_US
dc.contributor.departmentAnorganisch - Chemisches Institut der Technischen Universität München, Abteilung Analytische Chemie, Lichtenbergstr. 4, D - 8046 Garching, FRG.en_US
dc.identifier.journalBiosensors International Workshop 1987en_US
refterms.dateFOA2023-08-30T09:47:18Z


Files in this item

Thumbnail
Name:
Weiss267.pdf
Size:
1015.Kb
Format:
PDF
Description:
PDF

This item appears in the following Collection(s)

Show simple item record

Attribution-NonCommercial-ShareAlike 4.0 International
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-ShareAlike 4.0 International