SIMPLIFIED DETECTION OF BIOLOGICAL HALOGENATION AND DEHALOGENATION WITH A HALIDE SENSOR
Average rating
Cast your vote
You can rate an item by clicking the amount of stars they wish to award to this item.
When enough users have cast their vote on this item, the average rating will also be shown.
Star rating
Your vote was cast
Thank you for your feedback
Thank you for your feedback
Issue Date
1987Submitted date
2023-08-30
Metadata
Show full item recordAbstract
Flow injection analysis of halides with a halide selective electrode is a sensitive method to follow changes of halide concentrations which accompany biological halogenation and dehalogenation reactions. Less than 10 picomole chloride or 2 picomole bromide can be detected. At low halide concentrations the halide concentration and the potential change correlate linearily.Citation
Biosensors International Workshop 1987, 279 - 280Affiliation
FraunhoferInstitut für Grenzflächen- und Bioverfahrenstechnik Nobelstraße 12, 7000 Stuttgart 80, FRG *Institut für Mikrobiologie der Universität Hohenheim Garbenstr. 30, 7000 Stuttgart 70, FRGType
Book chapterconference paper
Language
enSeries/Report no.
GBF Monographs, Volume 10ISSN
0930-4320ISBN
0-89573-683-73-527-26801-4
Collections
The following license files are associated with this item:
- Creative Commons
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-ShareAlike 4.0 International