FLOW INJECTION IMMUNOANALYSIS (FIIA) FOR THE DETERMINATION OF PESTICIDES IN WATER
Average rating
Cast your vote
You can rate an item by clicking the amount of stars they wish to award to this item.
When enough users have cast their vote on this item, the average rating will also be shown.
Star rating
Your vote was cast
Thank you for your feedback
Thank you for your feedback
Issue Date
1989Submitted date
2024-01-16
Metadata
Show full item recordAbstract
The principle of a heterogeneous competitive enzyme immunoassay was applied to a flow injection system. This flow injection immunoassay (FIIA) can be run automatically. Antibodies to the triazine herbicide atrazin were immobilized on membranes. Antibody containing membranes were mounted in a cross-flow reactor and changed after each assay. In the assay atrazin competed with an atrazin-peroxidase conjugate for the antibody binding sites. The product of the enzyme reaction was measured fluorimetrically. One assay is completed within 15 minutes. Sensitivities of the FIIA proved similar to the ELISA and depended mainly on the properties of the antibodies and conjugation strategies. The potential and limitations of the FIIA system will be discussed.Citation
Biosensors : applications in medicine, environmental protection and process control, 307 - 312Affiliation
GBF - Gesellschaft für Biotechnologische Forschung, Abt. für Enzymtechnologie, Mascheroder Weg 1, D-3300 Braunschweig, FRGType
Book chapterconference paper
Language
enSeries/Report no.
GBF monographs ; Volume 13ISSN
0930-4320ISBN
35272803240895739550
Collections
The following license files are associated with this item:
- Creative Commons
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-ShareAlike 4.0 International