DEVELOPMENT OF AN ENZYMATIC MULTICHANNEL FLOW INJECTION ANALYSIS SYSTEM FOR MONITORING MAMMALIAN CELL FERMENTATIONS
Average rating
Cast your vote
You can rate an item by clicking the amount of stars they wish to award to this item.
When enough users have cast their vote on this item, the average rating will also be shown.
Star rating
Your vote was cast
Thank you for your feedback
Thank you for your feedback
Issue Date
1992Submitted date
2024-05-08
Metadata
Show full item recordAbstract
A multichannel flow injection analysis system described earlier /1/, is adapted to mammalian cell fermentations by increasing the sensitivity of the system. The analysis is based on the conversion of the compounds by immobilized dehydrogenases, resulting in the built up or consumption of NADH which is measured with a fluorescence- detector. The method presented here has a linear glucose determination range of 0.05 g/1 to 3.5 g/1 and a non-linear range for lactate of 0.05 g/1 to 2.7 g/l.Citation
Biosensors : fundamentals, technologies and applications, 349 - 352Affiliation
Institut für Biotechnologie, Forschungszentrum Jülich, Postfach 1913, D-5170 Jülich, Germany; Food and Bioprocess Engineering Group, Agricultural University Wageningen, P.O. Box 8129, NL-6700 EV, Wageningen, The NetherlandsType