ENZYME ELECTRODES FOR IN VIVO APPLICATION; KINETIC PROPERTIES; STERILIZATION, AND GEOMETRY
Average rating
Cast your vote
You can rate an item by clicking the amount of stars they wish to award to this item.
When enough users have cast their vote on this item, the average rating will also be shown.
Star rating
Your vote was cast
Thank you for your feedback
Thank you for your feedback
Issue Date
1992Submitted date
2024-05-08
Metadata
Show full item recordAbstract
Using glucose oxidase/hydrogen peroxide electrodes, covered with Cuprophane and sealed with cellulose acetate or polyurethane, there are three major difficulties in practical application of subcutaneous implantable glucose electrodes. These are the lack of knowledge concerning kinetic properties between different sensor preparations and their influence on the response characteristics in vivo, the suitable sterilization procedure, and the sensor geometry. Response times of the sensors in vitro were between 1 and 5 min (testgeie 3 of according to nonlinear regression analysis, NLRA), in dependence on qualities and thickness of the covering layer. The time constants T resulting from in vivo measurings subjected to NLRA at increases and decreases of glucose were 28+8 and 15+2 min, in blood, 26.5+5 and 18+2 min in plasma, and 53+10 and 2644 min in subcutaneous tissue. As a practicable methode to sterilize enzyme sensors, gamma irradation in the presence of hydrogen peroxide has been used at doses of 0.6 kGy + 0.1 % H202 for safe killing Pseudomonas aeruginosa. Increasing H202- concentration (at about 1 %), however, reduces the sensitivity by influencing the enzyme activity. Overcoming problems at sensors implantation site caused by the size of the sensor (d = 2 mm, 1 = 20 mm) most probably, a miniaturized dualelectrode was constructed, where the working electrode (Pt-anode, d < 0.5 mm) has been prepared as an enzyme electrode, only. This electrode arrangement has shown excellent electrochemical characteristics as well as in the pOz - and in the H202-polarogran.Citation
Biosensors : fundamentals, technologies and applications, 369 - 368Affiliation
Institute of Diabetes "Gerhardt Katsch", Karlsburg, Germany and Institute of Hygiene of Mecklenburg/Vorpommern, Greifswald, GermanyType
Book chapterconference paper
Language
enSeries/Report no.
GBF monographs ; Volume 17ISSN
0930-4320ISBN
15608122063527284370
Collections
The following license files are associated with this item:
- Creative Commons
Except where otherwise noted, this item's license is described as Attribution-NonCommercial-ShareAlike 4.0 International